18F, 68Ga, 89Zr Accreditation specifications
The EARL accreditation is based on quality control (QC) experiments that were described and published in the EANM guidelines in 2010. Meanwhile the field has moved forward and new developments, technologies and insights have become available. During the EANM Conference 2018 in Düsseldorf EARL announced the update of the EARL performance specifications. The current standard allows sites to benefit from state-of-the-art PET/CT technology and is based on a recent publication by EARL entitled “Feasibility of state-of-the-art PET/CT systems performance harmonisation”, by Kaalep et al. EJNMMI 2018. The impact of the new 18F standard 2 on the quantification of clinical studies has been described in “Quantitative implications of the updated EARL 2019 PET–CT performance standards”, by Kaalep et al. EJNMMI Physics 2019. EARL is dedicated towards improving and supporting the quality of nuclear medicine procedures.
Click here to review presentation on this EARL update, given by Prof. Dr. R. Boellaard.
Calibration QC acceptability criteria for 18F, 68Ga and 89Zr are set to +/-10% for each isotope independently.
SUV recovery coefficients for 18F standards 1 and 2 effective as of quarter 3 of 2020:
NEMA IEC phantom spheres |
18F standards 1 RCs |
18F standards 2 RCs |
||||
Diameter (mm) |
Volume (mL) |
max |
mean |
max* |
mean |
peak** |
37 |
26.52 |
0.95 – 1.16 |
0.76 – 0.89 |
1.05 – 1.29 |
0.85 – 1.00 |
0.90 – 1.10 |
28 |
11.49 |
0.91 – 1.13 |
0.72 – 0.85 |
1.01 – 1.26 |
0.82 – 0.97 |
0.90 – 1.10 |
22 |
5.57 |
0.83 – 1.09 |
0.63 – 0.78 |
1.01 – 1.32 |
0.80 – 0.99 |
0.90 – 1.10 |
17 |
2.57 |
0.73 – 1.01 |
0.57 – 0.73 |
1.00 – 1.38 |
0.76 – 0.97 |
0.75 – 0.99 |
13 |
1.15 |
0.59 – 0.85 |
0.44 – 0.60 |
0.85 – 1.22 |
0.63 – 0.86 |
0.45 – 0.70 |
10 |
0.52 |
0.34 – 0.57 |
0.27 – 0.43 |
0.52 – 0.88 |
0.39 – 0.61 |
0.27 – 0.41 |
*SUVmax for total body PET/CT scanners is under investigation **SUVpeak limits are under a revision
Recovery coefficient values are calculated based on:- RCSUVmean – 50% background corrected isocontour VOI
- RCSUVmax – the maximum voxel value included in VOI
- RCSUVpeak – spherical VOI with a diameter of 12 mm, positioned so to yield the highest uptake